Investigation of emission contamination in the transmission image of a dual modality computed mammotomography system (Englisch)
- Neue Suche nach: Crotty, Dominic J.
- Neue Suche nach: Brzymialkewicz, Caryl N.
- Neue Suche nach: McKinley, Randolph L.
- Neue Suche nach: Tornai, Martin P.
- Neue Suche nach: Crotty, Dominic J.
- Neue Suche nach: Brzymialkewicz, Caryl N.
- Neue Suche nach: McKinley, Randolph L.
- Neue Suche nach: Tornai, Martin P.
In:
Proc. SPIE
;
6142
; 614237
;
2006
-
ISBN:
-
ISSN:
- Aufsatz (Konferenz) / Elektronische Ressource
-
Titel:Investigation of emission contamination in the transmission image of a dual modality computed mammotomography system
-
Beteiligte:Crotty, Dominic J. ( Autor:in ) / Brzymialkewicz, Caryl N. ( Autor:in ) / McKinley, Randolph L. ( Autor:in ) / Tornai, Martin P. ( Autor:in )
-
Kongress:Medical Imaging 2006: Physics of Medical Imaging ; 2006 ; San Diego,California,United States
-
Erschienen in:Proc. SPIE ; 6142 ; 614237
-
Verlag:
- Neue Suche nach: SPIE
-
Erscheinungsdatum:28.02.2006
-
ISBN:
-
ISSN:
-
DOI:
-
Medientyp:Aufsatz (Konferenz)
-
Format:Elektronische Ressource
-
Sprache:Englisch
-
Datenquelle:
Inhaltsverzeichnis Konferenzband
Die Inhaltsverzeichnisse werden automatisch erzeugt und basieren auf den im Index des TIB-Portals verfügbaren Einzelnachweisen der enthaltenen Beiträge. Die Anzeige der Inhaltsverzeichnisse kann daher unvollständig oder lückenhaft sein.
- 61420A
-
Optimization of operating conditions in photon-counting multi-slit mammography based on Si-strip detectorsÅslund, Magnus / Cederström, Björn / Lundqvist, Mats / Danielsson, Mats et al. | 2006
- 61420B
-
Evaluation of a photon-counting breast tomosynthesis imaging systemMaidment, Andrew D. A. / Ullberg, Christer / Lindman, Karin / Adelöw, Leif / Egerström, Johan / Eklund, Mathias / Francke, Tom / Jordung, Ulf / Kristoffersson, Tomas / Lindqvist, Lars et al. | 2006
- 61420C
-
High-speed large angle mammography tomosynthesis systemEberhard, Jeffrey W. / Staudinger, Paul / Smolenski, Joe / Ding, Jason / Schmitz, Andrea / McCoy, Julie / Rumsey, Michael / Al-Khalidy, Abdulrahman / Ross, William / Landberg, Cynthia E. et al. | 2006
- 61420D
-
Quantification for contrast-enhanced digital breast tomosynthesisCarton, Ann-Katherine / Li, Jingjing / Albert, Michael / Chen, Sara / Maidment, Andrew D. A. et al. | 2006
- 61420E
-
Gaussian frequency blending algorithm with matrix inversion tomosynthesis (MITS) and filtered back projection (FBP) for better digital breast tomosynthesis reconstructionChen, Ying / Lo, Joseph Y. / Baker, Jay A. / Dobbins III, James T. et al. | 2006
- 61420F
-
Optimizing filtered backprojection reconstruction for a breast tomosynthesis prototype deviceMertelmeier, Thomas / Orman, Jasmina / Haerer, Wolfgang / Dudam, Mithun K. et al. | 2006
- 61420G
-
Motion artifacts from an inverse-geometry CT system with multiple detector arraysMazin, Samuel / Pelc, Norbert et al. | 2006
- 61420H
-
Design considerations in cardiac CTBruder, H. / Stierstorfer, K. / McCullough, C. / Raupach, R. / Petersilka, M. / Grasruck, M. / Suess, C. / Ohnesorge, B. / Flohr, T. et al. | 2006
- 61420I
-
Effect of heart rate on CT angiography using the enhanced cardiac model of the 4D NCATSegars, W. P. / Taguchi, K. / Fung, G. S. K. / Fishman, E. K. / Tsui, B. M. W. et al. | 2006
- 61420J
-
Toward time resolved 4D cardiac CT imaging with patient dose reduction: estimating the global heart motionTaguchi, Katsuyuki / Segars, W. Paul / Fung, George S. K. / Tsui, Benjamin M. W. et al. | 2006
- 61420K
-
Optimized time window for cardiac CT reconstruction based on Doppler tissue imaging (DTI) dataGuerra, Rui / Codreanu, Andre / Ponvianne, Yannick / Grosjean, Romain / Felblinger, Jacques et al. | 2006
- 61420L
-
ECG gated continuous circular cone-beam multi-cycle reconstruction for in-stent coronary artery imaging: a phantom studyvan Stevendaal, Udo / Koken, Peter / Begemann, Philipp G. C. / Koester, Ralf / Adam, Gerhard / Grass, Michael et al. | 2006
- 61420M
-
Impact of noise on image reconstruction for diffuse optical tomographyNielsen, Tim / Köhler, Thomas et al. | 2006
- 61420N
-
High-resolution multiphoton optical tomography of tissues: an in vitro and in vivo studyRiemann, Iris / Schenke-Layland, Katja / Ehlers, Alexander / Dimitrow, Enrico / Kaatz, Martin / Elsner, Peter / Martin, Sven / König, Karsten et al. | 2006
- 61420O
-
Generalized auto-calibrating technique for image reconstruction from sensitivity encoded MRI dataKholmovski, Eugene G. / Parker, Dennis L. et al. | 2006
- 61420P
-
Reducing temporal fluctuations in MRI with the multichannel method SENSEMoeller, Steen / Van de Moortele, Pierre-Francois / Goerke, Ute / Uğurbil, Kâmil et al. | 2006
- 61420Q
-
Description of statistical theory of magnetic resonance imaging (MRI)Lei, Tianhu et al. | 2006
- 61420R
-
Performance evaluation of dual-crystal APD-based detector modules for positron emission tomographyPepin, Catherine M. / Bérard, Philippe / Cadorette, Jules / Tétrault, Marc-André / Leroux, Jean-Daniel / Michaud, Jean-Baptiste / Robert, Stéfan / Dautet, Henri / Davies, Murray / Fontaine, Réjean et al. | 2006
- 61420S
-
High voltage protection in active matrix flat-panel imagersLehnert, Joerg / Zhao, Wei et al. | 2006
- 61420T
-
Low-noise pixel architecture for advanced diagnostic medical x-ray imaging applicationsIzadi, M. H. / Karim, K. S. / Nathan, A. / Rowlands, J. A. et al. | 2006
- 61420U
-
A new 2D-tiled detector for multislice CTLuhta, Randy / Chappo, Marc / Harwood, Brian / Mattson, Rod / Salk, Dave / Vrettos, Chris et al. | 2006
- 61420V
-
Novel CT detector based on an inorganic scintillator working in photon-counting modeBérard, Philippe / Riendeau, Joël / Pepin, Catherine M. / Rouleau, Daniel / Cadorette, Jules / Fontaine, Réjean / Lecomte, Roger et al. | 2006
- 61420W
-
Three-dimensional columnar CsI model for x-ray imaging system simulations using MANTIS: validating for noise, blur, and light outputBadano, Aldo / Kyprianou, Iacovos S. / Sempau, Josep et al. | 2006
- 61420X
-
Evaluation of noise and resolution properties of penalized-likelihood CT sinogram smoothing and restoration algorithmsLa Rivière, Patrick J. / Vargas, Phillip et al. | 2006
- 61420Y
-
Comparison of three sinogram restoration methodsForthmann, Peter / Köhler, Thomas / Defrise, Michel / La Rivière, Patrick J. et al. | 2006
- 61420Z
-
Investigation of image lag in a high-speed flat panel detector-based cone beam CT imaging systemZhang, Yan / Ning, Ruola / Conover, David et al. | 2006
- 61420Z
-
Investigation of image lag in a high-speed flat panel detector-based cone beam CT imaging system [6142-35]Zhang, Y. / Ning, R. / Conover, D. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 61421A
-
The impact of angular separation on the performance of biplane correlation imaging for lung nodule detectionNasab, Nariman Majdi / Samei, Ehsan et al. | 2006
- 61421B
-
Evaluation of a physical based approach of scattered radiation correction in cone beam CT with an anthropomorphic thorax phantomRinkel, Jean / Gerfault, Laurent / Estève, François / Dinten, Jean-Marc et al. | 2006
- 61421C
-
Scatter correction for cone-beam computed tomography using simulated object modelsBertram, Matthias / Wiegert, Jens / Rose, Georg et al. | 2006
- 61421D
-
Development of next generation digital flat panel catheterization system: design principles and validation methodologyBelanger, B. / Betraoui, F. / Dhawale, P. / Gopinath, P. / Tegzes, Pal / Vagvolgyi, B. et al. | 2006
- 61421E
-
High-performance dual-energy imaging with a flat-panel detector: imaging physics from blackboard to benchtop to bedsideSiewerdsen, J. H. / Shkumat, N. A. / Dhanantwari, A. C. / Williams, D. B. / Richard, S. / Daly, M. J. / Paul, N. S. / Moseley, D. J. / Jaffray, D. A. / Yorkston, J. et al. | 2006
- 61421F
-
The impact of cardiac gating on the detection of coronary calcifications in dual-energy chest radiography: a phantom studySabol, John M. / Liu, Ray / Saunders, Rowland / Markley, Jonathan / Moreno, Nery / Seamans, John / Wiese, Scott / Jabri, Kadri / Gilkeson, Robert C. et al. | 2006
- 61421G
-
Accuracy and precision of dual energy CT imaging for the quantification of tissue fat contentWalter, Deborah J. / Tkaczyk, Eric J. / Wu, Xiaoye et al. | 2006
- 61421H
-
Dual-energy imaging using a photon counting detector with electronic spectrum-splittingBornefalk, Hans / Lundqvist, Mats et al. | 2006
- 61421I
-
Multispectral single-scan lung imaging system: initial feasibility [6142-54]Besson, G. M. / Crocker, K. E. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 61421I
-
Multispectral single-scan lung imaging system: initial feasibilityBesson, Guy M. / Crocker, Kenneth E. et al. | 2006
- 61421J
-
Simulated and experimental technique optimization of dual-energy radiography: abdominal imaging applicationsSabol, John M. / Wheeldon, Samuel J. / Jabri, Kadri N. et al. | 2006
- 61421K
-
Light transport in trabecular bone: Monte Carlo simulation based on 3D triangle meshesMargallo-Balbás, Eduardo / French, Patrick J. / Wieringa, Peter A. et al. | 2006
- 61421L
-
Optimal spectra for indirect detector breast tomosynthesisGlick, Stephen J. / Gong, Xing et al. | 2006
- 61421M
-
Development of a computer-generated model for the coronary arterial tree based on multislice CT and morphometric dataFung, George S. K. / Segars, W. Paul / Taguchi, Katsuyuki / Fishman, Elliot K. / Tsui, Benjamin M. W. et al. | 2006
- 61421N
-
Monte Carlo simulation of x-ray scatter based on patient model from digital breast tomosynthesisLiu, Bob / Wu, Tao / Moore, Richard H. / Kopans, Daniel B. et al. | 2006
- 61421O
-
Breast cancer diagnosis using neutron stimulated emission computed tomography: dose and count requirementsFloyd Jr., Carey E. / Bender, Janelle E. / Harrawood, Brian / Sharma, Amy C. / Kapadia, Anuj / Tourassi, Georgia D. / Lo, Joseph Y. / Howell, Calvin et al. | 2006
- 61421P
-
Theoretical investigation of very high-quantum-efficiency segmented crystalline detectors for low-contrast visualization in megavoltage cone-beam CTWang, Yi / Antonuk, Larry E. / El-Mohri, Youcef / Sawant, Amit / Zhao, Qihua / Du, Hong / Li, Yixin et al. | 2006
- 61421Q
-
CMOS cassette for digital upgrade of film-based mammography systemsBaysal, Mehmet A. / Toker, Emre et al. | 2006
- 61421R
-
New light-amplifier-based detector designs for high spatial resolution and high sensitivity CBCT mammography and fluoroscopyRudin, Stephen / Kuhls, Andrew T. / Yadava, Girijesh K. / Josan, Gaurav C. / Wu, Ye / Chityala, Ravishankar N. / Rangwala, Hussain S. / Ionita, N. Ciprian / Hoffmann, Kenneth R. / Bednarek, Daniel R. et al. | 2006
- 61421S
-
Effect of ghosting on the modulation transfer function of amorphous selenium based flat panel detectorsChowdhury, Lamia / Tousignant, Olivier / DeCrescenzo, Giovanni / Gauthier, P. / Leboeuf, J. / Rowlands, J. A. et al. | 2006
- 61421T
-
Single-photon spatial and energy resolution enhancement of a columnar CsI(Tl)/EMCCD gamma-camera using maximum-likelihood estimationMiller, Brian W. / Barber, H. Bradford / Barrett, Harrison H. / Shestakova, Irina / Singh, Bipin / Nagarkar, Vivek V. et al. | 2006
- 61421U
-
A practical method for measuring the H matrix of digital x-ray and cone beam CT imaging systemsKyprianou, Iacovos S. / Badano, Aldo / Gallas, Brandon D. / Park, Subok / Myers, Kyle J. et al. | 2006
- 61421V
-
Experimental spectral measurements of heavy K-edge filtered beams for x-ray computed mammotomographyCrotty, Dominic J. / McKinley, Randolph L. / Tornai, Martin P. et al. | 2006
- 61421W
-
Preliminary validation of a new methodology for estimating dose reduction protocols in neonatal chest computed radiographsDon, Steven / Whiting, Bruce R. / Hildebolt, Charles F. / Sehnert, W. James / Ellinwood, Jacquelyn S. / Töpfer, Karin / Masoumzadeh, Parinaz / Kraus, Richard A. / Kronemer, Keith A. / Herman, Thomas et al. | 2006
- 61421X
-
A method to measure the temporal MTF to determine the DQE of fluoroscopy systemsFriedman, S. N. / Cunningham, I. A. et al. | 2006
- 61421Y
-
A novel method to characterize the MTF in 3D for computed mammotomographyMadhav, Priti / McKinley, Randolph L. / Samei, Ehsan / Bowsher, James E. / Tornai, Martin P. et al. | 2006
- 61421Z
-
Measurement of the spatial resolution of a clinical volumetric computed tomography scanner using a sphere phantomThornton, Michael M. / Flynn, Michael J. et al. | 2006
- 61421Z
-
Measurement of the spatial resolution of a clinical volumetric computed tomography scanner using a sphere phantom [6142-71]Thornton, M. M. / Flynn, M. J. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 61422A
-
A new reconstruction algorithm for Radon dataXu, Y. / Tischenko, O. / Hoeschen, C. et al. | 2006
- 61422B
-
Data pre-processing for quantification in tomography and radiography with a digital flat panel detectorRinkel, Jean / Gerfault, Laurent / Estève, François / Dinten, Jean-Marc et al. | 2006
- 61422C
-
Flat panel detector-based cone beam CT for dynamic imaging: system evaluationNing, Ruola / Conover, David / Yu, Yong / Zhang, Yan / Cai, Weixing / Yang, Dong / Lu, Xianghua et al. | 2006
- 61422D
-
Experimental comparison of cone beam CT (CBCT) reconstruction and multiview reconstruction (MVR) for microangiography (MA) detector systemPatel, Vikas / Kuhls, Andrew T. / Noël, Peter B. / Walczak, Alan / Ionita, Ciprian N. / Chityala, Ravishankar / Tranquebar, Rekha / Rangwala, Hussain S. / Kasodekar, Snehal S. / Hoffmann, Kenneth R. et al. | 2006
- 61422E
-
Geometry calibration phantom design for 3D imagingClaus, Bernhard E. H. et al. | 2006
- 61422F
-
Impact of CT detector pixel-to-pixel crosstalk on image qualityEngel, Klaus J. / Spies, Lothar / Vogtmeier, Gereon / Luhta, Randy et al. | 2006
- 61422G
-
Image quality improvement and exposure dose reduction with the combined use of X-ray modulation and Boost3DKazama, Masahiro / Tsukagoshi, Shinsuke / Okumura, Miwa et al. | 2006
- 61422H
-
Windmill artifacts analysis in MSCTAmir, O. / Sabo-Napadensky, I. et al. | 2006
- 61422I
-
Amplitude correlated 4-dimensional cone beam CTLu, Jun / Zhu, X. Ronald / Munro, Peter / Chi, Paichun M. / Mohan, Radhe / Pan, Tinsu et al. | 2006
- 61422J
-
Combination of CT scanning and fluoroscopy imaging on a flat-panel CT scannerGrasruck, M. / Gupta, R. / Reichardt, B. / Suess, Ch. / Schmidt, B. / Stierstorfer, K. / Popescu, S. / Brady, T. / Flohr, T. et al. | 2006
- 61422K
-
Comparison measurements of DQE for two flat panel detectors: fluoroscopic detector vs. cone beam CT detectorBetancourt Benítez, Ricardo / Ning, Ruola / Conover, David et al. | 2006
- 61422L
-
A new scanning device in CT with dose reduction potentialTischenko, Oleg / Xu, Yuan / Hoeschen, Christoph et al. | 2006
- 61422M
-
X-ray tube voltage and image quality in adult and pediatric CTHuda, W. / Ogden, K. M. / Scalzetti, E. M. / Lavallee, R. L. / Samei, E. et al. | 2006
- 61422N
-
Dual-energy contrast-enhanced digital mammography (DE-CEDM): optimization on digital subtraction with practical x-ray low/high-energy spectraChen, Biao / Jing, Zhenxue / Smith, Andrew P. / Parikh, Samir / Parisky, Yuri et al. | 2006
- 61422O
-
Dual-energy imaging using a digital scanned multi-slit system for mammography: evaluation of a differential beam filtering techniqueBornefalk, Hans / Hemmendorff, Magnus / Hjärn, Torbjörn et al. | 2006
- 61422P
-
Nonlinear dual-spectral image fusion for improving cone-beam-CT-based breast cancer diagnosisChen, Zikuan / Ning, Ruola / Conover, David / Willison, Kathleen et al. | 2006
- 61422R
-
An iteration algorithm in dual-energy x-ray imaging based on polychromatic physics modelTang, Shaojie / Mou, Xuanqin / Luo, Tao et al. | 2006
- 61422S
-
A computation method of dual-energy x-ray imagingMou, Xuanqin / Tang, Shaojie / Hong, Wei et al. | 2006
- 61422T
-
The line-noise-reduction for low dose x-ray fluoroscopy with the flat panel detectorNakamura, Tadashi / Ikeda, Shigeyuki / Suzuki, Katsumi / Takenouchi, Shinobu et al. | 2006
- 61422U
-
Light induced degradation in amorphous silicon photodiodes and implication for diagnostic medical imaging applicationKhodami, I. / Adachi, M.M. / Malhotra, M. / Taghibakhsh, F. / Rowlands, J.A. / Karim, K.S. / Kavanagh, K.L. et al. | 2006
- 61422V
-
Temperature coefficients and noise performance and studies for the back-illuminated arrays for medical imaging applicationsGoushcha, Alexander O. / Tabbert, Bernd / Goushcha, Ilja et al. | 2006
- 61422W
-
Novel x-ray image sensor using CsBr:Eu phosphor for computed radiographyNanto, H. / Takei, Y. / Nishimura, A. / Nakano, Y. / Shouji, T. / Yanagita, T. / Kasai, S. et al. | 2006
- 61422X
-
Transparent BaCl2:Eu2+glass-ceramic scintillatorChen, Gang / Johnson, Jacqueline / Schweizer, Stefan / Woodford, John / Newman, Peter / MacFarlane, Douglas et al. | 2006
- 61422Y
-
Energy-dependent scintillation intensity of fluorozirconate-based glass-ceramic x-ray detectorsSchweizer, Stefan / Henke, Bastian / Köneke, Stephanie / Johnson, Jacqueline A. / Chen, Gang / Woodford, John et al. | 2006
- 61422Z
-
Quantitative evaluation of mercuric iodide thick film for x-ray imaging device [6142-107]Kim, K.-J. / Kang, S.-S. / Park, J.-K. / Cho, S.-H. / Cha, B.-Y. / Shin, J.-W. / Nam, S.-H. / Kim, J.-H. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 61422Z
-
Quantitative evaluation of mercuric iodide thick film for x-ray imaging deviceKim, Kyung-Jin / Kang, Sang-Sik / Park, Ji-Koon / Cho, Sung-Ho / Cha, Byung-Youl / Shin, Jung-Wook / Nam, Sang-Hee / Kim, Jae-Hyung et al. | 2006
- 61423A
-
A Monte Carlo investigation on the impact of scattered radiation on mammographic resolution and noiseSaunders, Robert S. / Samei, Ehsan et al. | 2006
- 61423B
-
A comparison of the performance of new screen-film and digital mammography systemsMonnin, P. / Gutierrez, D. / Castella, C. / Lepori, D. / Verdun, F. R. et al. | 2006
- 61423C
-
Dynamic platform for moving organ imagingGrosjean, Romain / Guerra, Rui / Lorentz, Céline / Pasquier, Cédric / Vuissoz, Pierre-André / Claudon, Michel / Felblinger, Jacques et al. | 2006
- 61423D
-
Anode heel affect in thoracic radiology: a visual grading analysisMearon, T. / Brennan, P. C. et al. | 2006
- 61423E
-
Scatter correction using beam stop array algorithm for cone-beam CT breast imaging [6142-122]Cai, W. / Ning, R. / Conover, D. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 61423E
-
Scatter correction using beam stop array algorithm for cone-beam CT breast imagingCai, Weixing / Ning, Ruola / Conover, David et al. | 2006
- 61423F
-
Response of a CsI/amorphous-Si flat panel detector as function of incident x-ray angleTkaczyk, J. Eric / Claus, Bernhard / Gonzalez Trotter, Dinko / Eberhard, J. W. et al. | 2006
- 61423G
-
Dosimetric and image quality assessment of different acquisition protocols of a novel 64-slice CT scannerVite, Cristina / Mangini, Monica / Strocchi, Sabina / Novario, Raffaele / Tanzi, Fabio / Carrafiello, Gianpaolo / Conte, Leopoldo / Fugazzola, Carlo et al. | 2006
- 61423H
-
Novel NPS measurement method for medical liquid crystal display using periodic components subtraction techniqueIchikawa, Katsuhiro / Kodera, Yoshie / Horii, Akiko et al. | 2006
- 61423I
-
A method of accuracy evaluation of line spread function (LSF) and point spread function (PSF) measured in computed tomographyWada, Shinichi / Ohkubo, Masaki / Matsumoto, Toru / Nishizawa, Kanae et al. | 2006
- 61423J
-
Substrate effect on indirect digital radiography system performanceLubinsky, A. R. / Yip, Kwok L. / Trauernicht, David P. / Yorkston, John et al. | 2006
- 61423K
-
A robust x-ray tube spectra measuring method by attenuation dataYang, Ying / Mou, Xuanqin / Chen, Xi et al. | 2006
- 61423L
-
Choosing the right resolution for vessel visualization in MR angiographyVemuri, Prashanthi / Kholmovski, Eugene G. / Parker, Dennis L. et al. | 2006
- 61423M
-
New microangiography system development providing improved small vessel imaging, increased contrast-to-noise ratios, and multiview 3D reconstructionsKuhls, Andrew T. / Patel, Vikas / Ionita, Ciprian / Noël, Peter B. / Walczak, Alan M. / Rangwala, Hussain S. / Hoffmann, Kenneth R. / Rudin, Stephen et al. | 2006
- 61423N
-
Stationary x-ray source for tomographic medical imagingSchwoebel, Paul R. / Boone, John M. et al. | 2006
- 61423O
-
Novel SNR determination method in parallel MRIMiyati, Tosiaki / Imai, Hiroshi / Ogura, Akio / Doi, Tsukasa / Tsuchihashi, Toshio / Machida, Yoshio et al. | 2006
- 61423P
-
Fast parallel MRI reconstruction using B-spline approximation (PROBER)Petr, Jan / Kybic, Jan / Hlaváč, Václav / Müller, Sven / Bock, Michael et al. | 2006
- 61423Q
-
First attempt of the medical application of the refraction-based computed tomographyMaksimenko, Anton / Hashimoto, Eiko / Ando, Masami / Sugiyama, Hiroshi / Yuasa, Tetsuya et al. | 2006
- 61423S
-
Photo-bleaching compensation for autofocus algorithms in fluorescence microscope applicationsSoda, Paolo / Iannello, Giulio et al. | 2006
- 61423T
-
MR image reconstruction using the GPUSchiwietz, Thomas / Chang, Ti-chiun / Speier, Peter / Westermann, Rüdiger et al. | 2006
- 61423V
-
Spectral imaging of skin: experimental observations and analysesKerekes, John / Subramanian, Nithya / Kearney, Kevin / Schad, Nik et al. | 2006
- 61423X
-
Fat/water separation in a single MRI image with arbitrary phase shiftTisdall, M. Dylan / Atkins, M. Stella et al. | 2006
- 61423Y
-
Spectral imaging of near-surface oxygen saturationSubramanian, Nithya R. / Kerekes, John P. / Kearney, Kevin / Schad, Nik et al. | 2006
- 61423Z
-
Effects of collimator dependency and correction methods on I-123 SPECT images using x-ray-based attenuation mapYang, Yu-Wen / Chen, Jyh-Cheng / Chang, Chia-Jung / Cheng, Cheng-Yi / Huang, Wen-Sheng et al. | 2006
- 61423Z
-
Effects of collimator dependency and correction methods on I-123 SPECT images using x-ray-based attenuation map [6142-144]Yang, Y.-W. / Chen, J.-C. / Chang, C.-J. / Cheng, C.-Y. / Huang, W.-S. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 61424A
-
An inversion method for the exponential radon transform based on the harmonic analysis of the Euclidean motion groupYarman, Can Evren / Yazıcı, Birsen et al. | 2006
- 61424B
-
Elliptical extrapolation of truncated 2D CT projections using Helgason-Ludwig consistency conditionsVan Gompel, G. / Defrise, M. / Van Dyck, D. et al. | 2006
- 61424C
-
An inversion method for the cone-beam transformYarman, Can Evren / Yazıcı, Birsen et al. | 2006
- 61424D
-
A new approach of 3D SPECT reconstruction for near-field coded aperture imagingHong, Baoming / Mu, Zhiping / Liu, Yi-Hwa et al. | 2006
- 61424E
-
Noise reduction for low-dose helical CT by 3D penalized weighted least-squares sinogram smoothingWang, Jing / Li, Tianfang / Lu, Hongbing / Liang, Zhengrong et al. | 2006
- 61424F
-
Quarantine MAP reconstruction of PET/CT data using dual priorsChan, Chung / Meikle, Steven / Fulton, Roger / Cai, Tom W. / Feng, David D. et al. | 2006
- 61424G
-
Computer simulation of FDK reconstruction with the in-line holographic projection dataCai, Weixing / Ning, Ruola / Yang, Dong et al. | 2006
- 61424H
-
Reconstruction implementation based on a flat panel detector cone beam breast CTYang, Dong / Ning, Ruola / Conover, David / Yu, Yong et al. | 2006
- 61424I
-
Interpolation method for the Mojette transformServières, Myriam / Normand, Nicolas / Guédon, Jean-Pierre et al. | 2006
- 61424J
-
Image reconstruction in cone-beam CT with a spherical detector using the BPF algorithmZuo, Nianming / Zou, Yu / Jiang, Tianzi / Pan, Xiaochuan et al. | 2006
- 61424K
-
Effect of random coincidences for quantitative cardiac PET studies using 3D oxygen-15 water scansBouchareb, Y. / Thielemans, K. / Spinks, T. / Rimoldi, O. / Camici, P. G. et al. | 2006
- 61424L
-
Acceleration of fluoro-CT reconstruction for a mobile C-arm on GPU and FPGA hardware: a simulation studyXue, Xinwei / Cheryauka, Arvi / Tubbs, David et al. | 2006
- 61424M
-
Comparison of full-scan and half-scan for cone beam breast CT imagingChen, Lingyun / Shaw, Chris C. / Lai, Chao-jen / Altunbas, Mustafa C. / Wang, Tianpeng / Tu, Shu-ju / Liu, Xinming et al. | 2006
- 61424N
-
A backprojection filtered image reconstruction algorithm for circular cone-beam CTLi, Liang / Chen, Zhiqiang / Zhang, Li / Xing, Yuxiang / Kang, Kejun et al. | 2006
- 61424O
-
Implementation of strip-area system model for fan-beam collimator SPECT reconstructionYe, Hongwei / Krol, Andrzej / Feiglin, David H. / Lipson, Edward D. / Lee, Wei / Coman, Ioana L. et al. | 2006
- 61424P
-
Accelerating 3D cone beam T-FDK algorithm on commodity PC graphics hardwareDai, Zhisheng / Chen, Zhiqiang / Xing, Yuxiang / Zhang, Li et al. | 2006
- 61424R
-
Radial intensity projection for lumen: application to CT angiographic imagingKokubun, Hiroto / Miyazaki, Osamu / Hayashi, Hiromitsu et al. | 2006
- 61424S
-
Evaluation of an algorithm for simultaneous reconstruction of activity and attenuation in 3D PETSoto, Juan L. P. / Furuie, Sergio S. et al. | 2006
- 61424T
-
Image reconstruction from discrete Chebyshev moments via formation of lookup tablesBayraktar, Bulent / Bernas, Tytus / Robinson, J. Paul / Rajwa, Bartek et al. | 2006
- 61424U
-
An approximate cone beam reconstruction algorithm for gantry-tilted CTYan, Ming / Zhang, Cishen / Liang, Hongzhu et al. | 2006
- 61424V
-
Implementing an iterative reconstruction algorithm for digital breast tomosynthesis on graphics processing hardwareGoddard, Iain / Wu, Tao / Thieret, Scott / Berman, Ari / Bartsch, Hauke et al. | 2006
- 61424W
-
A correction method for unfunctional cell distortion using orthogonal polynomialsTang, Shaojie / Mou, Xuanqin / Yan, Hao et al. | 2006
- 61424X
-
Implementation of a spiral CT backprojection algorithm on the cell broadband engine processorBockenbach, Olivier / Goddard, Iain / Schuberth, Sebastian / Seebass, Martin et al. | 2006
- 61424Y
-
Cone-beam CT reconstruction using a nonlinear weighted filtered backprojection from half-scan dataGomi, Tsutomu / Koshida, Kichrou / Miyati, Tosiaki et al. | 2006
- 61424Z
-
The effect of detector resolution for quantitative analysis of neutron stimulated emission computed tomographyBender, Janelle E. / Floyd, Carey E. / Harrawood, Brian P. / Kapadia, Anuj J. / Sharma, Amy C. / Jesneck, Jonathan L. et al. | 2006
- 61424Z
-
The effect of detector resolution for quantitative analysis of neutron stimulated emission computed tomography [6142-180]Bender, J. E. / Floyd, C. E. / Harrawood, B. P. / Kapadia, A. J. / Sharmc, A. C. / Jesneck, J. L. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 61425A
-
Low-contrast lesion detection in tomosynthetic breast imaging using a realistic breast phantomZhou, Lili / Oldan, Jorge / Fisher, Paul / Gindi, Gene et al. | 2006
- 61425B
-
Selective photon counter for digital x-ray mammography tomosynthesisGoldan, Amir H. / Karim, Karim S. / Rowlands, J. A. et al. | 2006
- 61425C
-
Preliminary assessment of the temporal subtraction of tomosynthesis images for improved detection of pulmonary nodulesLi, Christina M. / Dobbins, James T. et al. | 2006
- 61425D
-
Measurement of slice thickness and in-plane resolution on radiographic tomosynthesis system using modulation transfer function (MTF)Li, Baojun / Saunders, Rowland / Uppaluri, Renuka et al. | 2006
- 61425E
-
Optimal acquisition techniques for digital breast tomosynthesis screening [6142-196]Wu, T. / Liu, B. / Moore, R. / Kopans, D. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 61425E
-
Optimal acquisition techniques for digital breast tomosynthesis screeningWu, Tao / Liu, Bob / Moore, Richard / Kopans, Daniel et al. | 2006
- 614203
-
Low cost digital radiographic imaging systems: the x-ray light valve [6142-03]Rowlands, J. A. / Webster, C. A. / Koprinarov, I. / Oakham, P. / Schad, K. C. / Szeto, T. / Germann, S. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614203
-
Low cost digital radiographic imaging systems: the x-ray light valveRowlands, J. A. / Webster, Christie A. / Koprinarov, Ivaylo / Oakham, Peter / Schad, Kelly C. / Szeto, Timothy / Germann, Stephen et al. | 2006
- 614204
-
A multi-beam x-ray imaging system based on carbon nanotube field emitters [6142-04]Zhang, J. / Yang, G. / Lee, Y. Z. / Cheng, Y. / Gao, B. / Qiu, Q. / Lu, J. P. / Zhou, O. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614204
-
A multi-beam x-ray imaging system based on carbon nanotube field emittersZhang, J. / Yang, G. / Lee, Y. Z. / Cheng, Y. / Gao, B. / Qiu, Q. / Lu, J. P. / Zhou, O. et al. | 2006
- 614205
-
Applying the European protocol for the quality control of the physical and technical aspects of mammography screening threshold contrast visibility assessment to digital systemsVan Metter, Richard / Heath, Michael / Fletcher-Heath, Lynn et al. | 2006
- 614205
-
Applying the European protocol for the quality control of the physical and technical aspects of mammography screening threshold contrast visibility assessment to digital systems [6142-108]Van Metter, R. / Heath, M. / Fletcher-Heath, L. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614206
-
Comparison of software and human observers in reading images of the CDMAM test object to assess digital mammography systems [6142-06]Young, K. C. / Cook, J. J. H. / Oduko, J. M. / Bosmans, H. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614206
-
Comparison of software and human observers in reading images of the CDMAM test object to assess digital mammography systemsYoung, Kenneth C. / Cook, James J. H. / Oduko, Jennifer M. / Bosmans, Hilde et al. | 2006
- 614207
-
Anatomically adaptable automatic exposure control (AEC) for amorphous selenium (a-Se) full field digital mammography (FFDM) system [6142-07]Varjonen, M. / Strommer, P. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614207
-
Anatomically adaptable automatic exposure control (AEC) for amorphous selenium (a-Se) full field digital mammography (FFDM) systemVarjonen, Mari / Strömmer, Pekka et al. | 2006
- 614208
-
Preliminary investigation of dose for a dedicated mammotomography system [6142-08]McKinley, R. L. / Tornai, M. P. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614208
-
Preliminary investigation of dose for a dedicated mammotomography systemMcKinley, Randolph L. / Tornai, Martin P. et al. | 2006
- 614209
-
Comparison of polychromatic and monochromatic x-rays for imagingHoheisel, M. / Bernhardt, P. / Lawaczeck, R. / Pietsch, H. et al. | 2006
- 614209
-
Comparison of polychromatic and monochromatic x-rays for imaging [6142-09]Hoheisel, M. / Bernhardt, P. / Lawaczeck, R. / Pietsch, H. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614210
-
Design and development of C-arm based cone-beam CT for image-guided interventions: initial resultsChen, Guang-Hong / Zambelli, Joseph / Nett, Brian E. / Supanich, Mark / Riddell, Cyril / Belanger, Barry / Mistretta, Charles A. et al. | 2006
- 614210
-
Design and development of C-arm based cone-beam CT for image-guided interventions: initial results [6142-36]Chen, G.-H. / Zambelli, J. / Nett, B. E. / Supanich, M. / Riddell, C. / Belanger, B. / Mistretta, C. A. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614211
-
A novel cone beam breast CT scanner: preliminary system evaluation [6142-37]Ning, R. / Conover, D. / Yu, Y. / Zhang, Y. / Cai, W. / Lu, X. / Betancourt-Benitez, R. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614211
-
A novel cone beam breast CT scanner: preliminary system evaluationNing, Ruola / Conover, David / Yu, Yong / Zhang, Yan / Cai, Weixing / Lu, Xianghua / Betancourt-Benitez, Richardo et al. | 2006
- 614212
-
Guidance for cone-beam CT design: tradeoff between view sampling rate and completeness of scanning trajectories [6142-38]Chen, G.-H. / Siewerdsen, J. H. / Leng, S. / Moseley, D. / Nett, B. E. / Hsieh, J. / Jaffray, D. / Mistretta, C. A. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614212
-
Guidance for cone-beam CT design: tradeoff between view sampling rate and completeness of scanning trajectoriesChen, Guang-Hong / Siewerdsen, Jeffrey H. / Leng, Shuai / Moseley, Douglas / Nett, Brian E. / Hsieh, Jiang / Jaffray, David / Mistretta, C. A. et al. | 2006
- 614213
-
Application of x-ray computed tomography based on the refraction contrast to biomedicineHashimoto, Eiko / Maksimenko, Anton / Sugiyama, Hiroshi / Hyodo, Kazuyuki / Shimao, Daisuke / Yuasa, Tetsuya / Nishino, Yoshinori / Ishikawa, Tetsuya / Mori, Koichi / Arai, Yoshinori et al. | 2006
- 614213
-
Application of x-ray computed tomography based on the refraction contrast to biomedicine [6142-39]Hashimoto, E. / Maksimenko, A. / Sugiyama, H. / Hyodo, K. / Shimao, D. / Yuasa, T. / Nishino, Y. / Ishikawa, T. / Mori, K. / Arai, Y. et al. | 2006
- 614214
-
3D cryo-section/imaging of blood vessel lesions for validation of MRI dataSalvado, Olivier / Roy, Debashish / Heinzel, Meredith / McKinley, Eliot / Wilson, David et al. | 2006
- 614214
-
3D cryo-section/imaging of blood vessel lesions for validation of MRI data [6142-40]Salvado, O. / Roy, D. / Heinzel, M. / McKinley, E. / Wilson, D. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614215
-
Performance of different reflectance and diffuse optical imaging tomographic approaches in fluorescence molecular imaging of small animalsDinten, Jean-Marc / Petié, Philippe / da Silva, Anabela / Boutet, Jérôme / Koenig, Anne / Hervé, Lionel / Berger, Michel / Laidevant, Aurélie / Rizo, Philippe et al. | 2006
- 614215
-
Performance of different reflectance and diffuse optical imaging tomographic approaches in fluorescence molecular imaging of small animals [6142-41]Dinten, J.-M. / Petie, P. / da Silva, A. / Boutet, J. / Koenig, A. / Herve, L. / Berger, M. / Laidevant, A. / Rizo, P. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614216
-
Image quality assessment of a pre-clinical flat-panel volumetric micro-CT scanner [6142-42]Du, L. Y. / Lee, T. / Holdsworth, D. W. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614216
-
Image quality assessment of a pre-clinical flat-panel volumetric micro-CT scannerDu, Louise Y. / Lee, Ting-Yim / Holdsworth, David W. et al. | 2006
- 614217
-
Rotating slat collimator design for high-energy near-field imagingSharma, Amy / Floyd, Carey / Harrawood, Brian / Tourassi, Georgia / Kapadia, Anuj / Bender, Janelle / Lo, Joseph / Howell, Calvin et al. | 2006
- 614217
-
Rotating slat collimator design for high-energy near-field imaging [6142-43]Sharma, A. / Floyd, C. / Harrawood, B. / Tourassi, G. / Kapadia, A. / Bender, J. / Lo, J. / Howell, C. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614218
-
Regional exposure management with spatial x-ray gating [6142-44]Yatsenko, D. V. / Anderton, R. L. / Sikorski, K. / Kirby, R. M. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614218
-
Regional exposure management with spatial x-ray gatingYatsenko, Dimitri V. / Anderton, R. Larry / Sikorski, Kris / Kirby, Robert M. et al. | 2006
- 614219
-
Applying a proposed definition for receptor dose to digital projection imagesVan Metter, Richard / Yorkston, John et al. | 2006
- 614219
-
Applying a proposed definition for receptor dose to digital projection images [6142-45]Van Metter, R. / Yorkston, J. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614220
-
Extension of the reconstruction field-of-view using sinogram decomposition [6142-72]Zamyatin, A. A. / Silver, M. D. / Nakanishi, S. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614220
-
Extension of the reconstruction field-of-view using sinogram decompositionZamyatin, Alexander A. / Silver, Michael D. / Nakanishi, Satoru et al. | 2006
- 614221
-
Suppression of motion-induced streak artifacts along chords in fan-beam BPF-reconstructions of motion-contaminated projection dataKing, Martin / Xia, Dan / Yu, Lifeng / Pan, Xiaochuan / Giger, Maryellen et al. | 2006
- 614221
-
Suppression of motion-induced streak artifacts along chords in fan-beam BPF-reconstructions of motion-contaminated projection data [6142-73]King, M. / Xia, D. / Yu, L. / Pan, X. / Giger, M. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614222
-
Reconstructions from parallel- and fan-beam data with truncationsXia, Dan / Zou, Yu / Pan, Xiaochuan et al. | 2006
- 614222
-
Reconstructions from parallel- and fan-beam data with truncations [6142-74]Xia, D. / Zou, Y. / Pan, X. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614223
-
Iterative reconstruction of a region of interest for transmission tomographyZiegler, Andy / Nielsen, Tim / Grass, Michael et al. | 2006
- 614223
-
Iterative reconstruction of a region of interest for transmission tomography [6142-75]Ziegler, A. / Nielsen, T. / Grass, M. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614224
-
An iterative algorithm for soft tissue reconstruction from truncated flat panel projections [6142-76]Langan, D. / Claus, B. / Edic, P. / Vaillant, R. / De Man, B. / Basu, S. / Iatrou, M. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614224
-
An iterative algorithm for soft tissue reconstruction from truncated flat panel projectionsLangan, D. / Claus, B. / Edic, P. / Vaillant, R. / De Man, B. / Basu, S. / Iatrou, M. et al. | 2006
- 614225
-
Conjugate backprojection approach for cone beam artifact reduction [6142-77]Hsieh, J. / Tang, X. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614225
-
Conjugate backprojection approach for cone beam artifact reductionHsieh, Jiang / Tang, Xiangyang et al. | 2006
- 614226
-
Cone-beam local tomography image reconstruction on chords [6142-78]Anastasio, M. A. / Sidky, E. Y. / Zou, Y. / Pan, X. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614226
-
Cone-beam local tomography image reconstruction on chordsAnastasio, Mark A. / Sidky, E. Y. / Zou, Y. / Pan, X. et al. | 2006
- 614227
-
3D-weighted cone beam filtered backprojection (CB-FBP) algorithm for image reconstruction at low-helical pitches to improve noise characteristics and dose efficiencyTang, Xiangyang / Hsieh, Jiang / Nilsen, Roy A. et al. | 2006
- 614227
-
3D-weighted cone beam filtered backprojection (CB-FBP) algorithm for image reconstruction at low-helical pitches to improve noise characteristics and dose efficiency [6142-79]Tang, X. / Hsieh, J. / Nilsen, R. A. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614228
-
ROI-image reconstruction for a circle-arc trajectory [6142-80]Zou, Y. / Xia, D. / Pan, X. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614228
-
ROI-image reconstruction for a circle-arc trajectoryZou, Yu / Xia, Dan / Pan, Xiaochuan et al. | 2006
- 614229
-
Accurate image reconstruction in CT from projection data taken at few-views [6142-81]Sidky, E. Y. / Kao, C.-M. / Pan, X. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614229
-
Accurate image reconstruction in CT from projection data taken at few-viewsSidky, Emil Y. / Kao, Chien-Min / Pan, Xiaochuan et al. | 2006
- 614230
-
Investigation of diagnostic and image quality attributes: comparison of screen-film to CR mammographyFletcher-Heath, Lynn / Richards, Anne / Ryan-Kron, Susan et al. | 2006
- 614230
-
Investigation of diagnostic and image quality attributes: comparison of screen-film to CR mammography [6142-05]Fletcher-Heath, L. / Richards, A. / Ryan-Kron, S. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614231
-
Comparison of a-Se direct-conversion and CsI(Tl) indirect-conversion flat-panel digital detectors: a clinical assessment of image quality for general radiography applicationsBarski, Lori L. / Wang, Xiaohui / Wandtke, John / Waldman, David / Davis, Delphine / Foos, David H. / Dupin, Michael / Huang, Weidong / Yorkston, John et al. | 2006
- 614231
-
Comparison of a-Se direct-conversion and CsI(TI) indirect-conversion flat-panel digital detectors: a clinical assessment of image quality for general radiography applications (Honorable Mention Poster Award) [6142-109]Barski, L. L. / Wang, X. / Wandtke, J. / Waldman, D. / Davis, D. / Foos, D. H. / Dupin, M. / Huang, W. / Yorkston, J. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614232
-
Novel features of the x-ray scatter profile that are not modeled by convolution of the primaryTkaczyk, J. Eric / Trousset, Yves / Walter, Deborah / Du, Yanfeng / Thompson, Richard A. / Harrison, Daniel et al. | 2006
- 614232
-
Novel features of the x-ray scatter profile that are not modeled by convolution of the primary [6142-110]Tkaczyk, J. E. / Trousset, Y. / Walter, D. / Du, Y. / Thompson, R. A. / Harrison, D. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614233
-
Effects of radiation dose level on calcification visibility in cone beam breast CT: a preliminary studyLai, Chao-Jen / Shaw, Chris C. / Altunbas, Mustafa C. / Meng, Yang / Chen, Lingyun / Tu, Shu-Ju / Wang, Tianpeng / Liu, Xinming / Yang, Wei T. / Whitman, Gary J. et al. | 2006
- 614233
-
Effects of radiation dose level on calcification visibility in cone beam breast CT: a preliminary study [6142-111]Lai, C.-J. / Shaw, C. C. / Altunbas, M. C. / Meng, Y. / Chen, L. / Tu, S.-J. / Wang, T. / Liu, X. / Yang, W. T. / Whitman, G. J. et al. | 2006
- 614234
-
An accurate scatter measurement and correction technique for cone beam breast CT imaging using scanning sampled measurement (SSM)technique [6142-112]Liu, X. / Shaw, C. C. / Wang, T. / Chen, L. / Altunbas, M. C. / Kappadath, S. C. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614234
-
An accurate scatter measurement and correction technique for cone beam breast CT imaging using scanning sampled measurement (SSM)techniqueLiu, Xinming / Shaw, Chris C. / Wang, Tianpeng / Chen, Lingyun / Altunbas, Mustafa C. / Kappadath, S. Cheenu et al. | 2006
- 614235
-
Scattered radiation in flat-detector based cone-beam CT: analysis of voxelized patient simulationsWiegert, Jens / Bertram, Matthias et al. | 2006
- 614235
-
Scattered radiation in flat-detector based cone-beam CT: analysis of voxelized patient simulations [6142-113]Wiegert, J. / Bertram, M. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614236
-
Effects of scattered radiation and beam quality on low contrast performance in cone beam breast CT [6142-114]Altunbas, M. C. / Shaw, C. / Chen, L. / Wang, T. / Tu, S. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614236
-
Effects of scattered radiation and beam quality on low contrast performance in cone beam breast CTAltunbas, M. Cem / Shaw, Chris / Chen, Lingyun / Wang, Tianpeng / Tu, Shuju et al. | 2006
- 614237
-
Investigation of emission contamination in the transmission image of a dual modality computed mammotomography systemCrotty, Dominic J. / Brzymialkewicz, Caryl N. / McKinley, Randolph L. / Tornai, Martin P. et al. | 2006
- 614237
-
Investigation of emission contamination in the transmission image of a dual modality computed mammotomography system [6142-115]Crotty, D. J. / Brzymialkewicz, C. N. / McKinley, R. L. / Tornai, M. P. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614238
-
Comparison of scatter correction methods for CBCT [6142-116]Suri, R. E. / Virshup, G. / Zurkirchen, L. / Kaissl, W. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614238
-
Comparison of scatter correction methods for CBCTSuri, Roland E. / Virshup, Gary / Zurkirchen, Luis / Kaissl, Wolfgang et al. | 2006
- 614239
-
Calculation of images from an anthropomorphic chest phantom using Monte Carlo methodsUllman, Gustaf / Malusek, Alexandr / Sandborg, Michael / Dance, David R. / Alm Carlsson, Gudrun et al. | 2006
- 614239
-
Calculation of images from an anthropomorphic chest phantom using Monte Carlo methods [6142-117]Ullman, G. / Malusek, A. / Sandborg, M. / Dance, D. R. / Carlsson, G. A. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614240
-
Discrete tomography from micro-CT data: application to the mouse trabecular bone structure [6142-145]Batenburg, K. J. / Sijbers, J. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614240
-
Discrete tomography from micro-CT data: application to the mouse trabecular bone structureBatenburg, K. J. / Sijbers, J. et al. | 2006
- 614241
-
Development of a K-edge micro CT for the study of tumor angiogenesis in small animals [6142-146]Baldazzi, G. / Boilini, D. / Gambaccini, M. / Golfieri, R. / Orsola-Malpighi, S. / Lollini, P. L. / Margotti, A. / Masetti, S. / Nicoletti, G. / Pancaldi, G. et al. | 2006
- 614241
-
Development of a K-edge micro CT for the study of tumor angiogenesis in small animalsBaldazzi, G. / Bollini, D. / Gambaccini, M. / Golfieri, R. / Lollini, P. L. / Margotti, A. / Masetti, S. / Nicoletti, G. / Pancaldi, G. / Roma, L. et al. | 2006
- 614244
-
Improved diagnostics using polarization imaging and artificial neural networksKlimach, Uwe / Zhao, Hongzhi / Chen, Qiushui / Zou, Yingyin Kevin / Wang, Yue / Xuan, Jianhua et al. | 2006
- 614244
-
Improved diagnostics using polarization imaging and artificial neural networks [6142-149]Klimach, U. / Zhao, H. / Chen, Q. / Zou, Y. K. / Wang, Y. / Xuan, J. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614245
-
Comparison of two detector systems for cone beam CT small animal imaging: a preliminary study [6142-150]Meng, Y. / Shaw, C. C. / Liu, X. / Altunbas, M. C. / Wang, T. / Chen, L. / Tu, S.-J. / Kappadath, S. C. / Lai, C.-J. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614245
-
Comparison of two detector systems for cone beam CT small animal imaging: a preliminary studyMeng, Yang / Shaw, Chris C. / Liu, Xinming / Altunbas, Mustafa C. / Wang, Tianpeng / Chen, Lingyun / Tu, Shu-Ju / Kappadath, S. Cheeenu / Lai, Chao-Jen et al. | 2006
- 614246
-
Maximum likelihood reconstruction of circular cone-beam CT data [6142-151]Ziegler, A. / Kohler, T. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614246
-
Maximum likelihood reconstruction of circular cone-beam CT dataZiegler, Andy / Köhler, Thomas et al. | 2006
- 614247
-
Penalized weighted least-squares approach for low-dose x-ray computed tomography [6142-152]Wang, J. / Li, T. / Lu, H. / Liang, Z. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614247
-
Penalized weighted least-squares approach for low-dose x-ray computed tomographyWang, Jing / Li, Tianfang / Lu, Hongbing / Liang, Zhengrong et al. | 2006
- 614248
-
An extended half-scan Feldkamp-type CT reconstruction [6142-153]Liang, H. / Zhang, C. / Yan, M. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614248
-
An extended half-scan Feldkamp-type CT reconstructionLiang, Hongzhu / Zhang, Cishen / Yan, Ming et al. | 2006
- 614249
-
Tomosynthesis reconstruction using the simultaneous algebraic reconstruction technique (SART) on breast phantom data [6142-154]Zhang, Y. / Chan, H.-P. / Sahiner, B. / Wei, J. / Goodsitt, M. M. / Hadjiiski, L. M. / Ge, J. / Zhou, C. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614249
-
Tomosynthesis reconstruction using the simultaneous algebraic reconstruction technique (SART) on breast phantom dataZhang, Yiheng / Chan, Heang-Ping / Sahiner, Berkman / Wei, Jun / Goodsitt, Mitchell M. / Hadjiiski, Lubomir M. / Ge, Jun / Zhou, Chuan et al. | 2006
- 614250
-
Image quality simulation and verification of x-ray volume imaging systemsKroon, Han / Schoumans, Nicole / Snoeren, Ruud et al. | 2006
- 614250
-
Image quality simulation and verification of x-ray volume imaging systems [6142-181]Kroon, H. / Schoumans, N. / Snoeren, R. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614252
-
Correction of inter-crystal scatter effect in iterative image reconstruction of the jPET-D4 [6142-183]Lam, C. F. / Hagiwara, N. / Obi, T. / Yamaguchi, M. / Yamaya, T. / Murayama, H. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614252
-
Correction of inter-crystal scatter effect in iterative image reconstruction of the jPET-D4Lam, Chih Fung / Hagiwara, Naoki / Obi, Takashi / Yamaguchi, Masahiro / Yamaya, Taiga / Murayama, Hideo et al. | 2006
- 614253
-
A novel model of the geometric and detector response for limited angular sampling pinhole SPECTWietholt, Christian / Hsiao, Ing-Tsung / Clough, Anne V. / Chen, Chin-Tu et al. | 2006
- 614253
-
A novel model of the geometric and detector response for limited angular sampling pinhole SPECT [6142-185]Wietholt, C. / Hsiao, I.-T. / Clough, A. V. / Chen, C.-T. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614254
-
Scatter correction for kilovoltage cone-beam computed tomography (CBCT) images using Monte Carlo simulations [6142-186]Jarry, G. / Graham, S. A. / Jaffray, D. A. / Moseley, D. J. / Verhaegen, F. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614254
-
Scatter correction for kilovoltage cone-beam computed tomography (CBCT) images using Monte Carlo simulationsJarry, G. / Graham, S. A. / Jaffray, D. A. / Moseley, D. J. / Verhaegen, F. et al. | 2006
- 614255
-
Discrete and continuous description of a three-dimensional scene for quality control of radiotherapy treatment planning systems [6142-187]Denis, E. / Guedon, J. / Beaumont, S. / Normand, N. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614255
-
Discrete and continuous description of a three-dimensional scene for quality control of radiotherapy treatment planning systemsDenis, Eloise / Guédon, JeanPierre / Beaumont, Stéphane / Normand, Nicolas et al. | 2006
- 614256
-
Modeling of realistic raw data for image reconstruction: quantifying scattering noise in different CT geometries [6142-188]Schlatti, H. / Tischenko, O. / Hoeschen, C. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614256
-
Modeling of realistic raw data for image reconstruction: quantifying scattering noise in different CT geometriesSchlattl, H. / Tischenko, O. / Hoeschen, C. et al. | 2006
- 614257
-
Monte Carlo simulation for the hybrid detector designKim, So Yeong / Park, Ji Koon / Kang, Sang Sik / Cha, Byung Youl / Cho, Sung Ho / Jeong, Eun Seon / Nam, Sang Hee / Lee, Hyung Won / Choi, Heung Kook et al. | 2006
- 614257
-
Monte Carlo simulation for the hybrid detector design [6142-189]Kim, S. Y. / Park, J. K. / Kang, S. S. / Cha, B. Y. / Cho, S. H. / Jeong, E. S. / Nam, S. H. / Lee, H. W. / Choi, H. K. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614258
-
Characterization of point spread function in linear digital tomosynthesis: a simulation study [6142-190]Avinash, G. / Israni, K. / Li, B. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614258
-
Characterization of point spread function in linear digital tomosynthesis: a simulation studyAvinash, Gopal / Israni, Kenny / Li, Baojun et al. | 2006
- 614259
-
Noise power spectrum analysis for several digital breast tomosynthesis reconstruction algorithms [6142-191]Chen, Y. / Lo, J. Y. / Dobbins, J. T. / SPIE-- the International Society for Optical Engineering et al. | 2006
- 614259
-
Noise power spectrum analysis for several digital breast tomosynthesis reconstruction algorithmsChen, Ying / Lo, Joseph Y. / Dobbins, James T. et al. | 2006
-
Gaussian frequency blending algorithm with matrix inversion tomosynthesis (MITS) and filtered back projection (FBP) for better digital breast tomosynthesis reconstruction [6142-14]Chen, Y. / Lo, J. Y. / Baker, J. A. / Dobbins, J. T. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Evaluation of noise and resolution properties of penalized-likelihood CT sinogram smoothing and restoration algorithms [6142-33]La Riviere, P. J. / Vargas, P. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Novel CT detector based on an inorganic scintillator working in photon-counting mode [6142-31]Berard, P. / Riendeau, J. / Pepin, C. M. / Rouleau, D. / Cadorette, J. / Fontaine, R. / Lecomte, R. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
The impact of angular separation on the performance of biplane correlation imaging for lung nodule detection [6142-46]Nasab, N. M. / Samei, E. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Accuracy and precision of dual energy CT imaging for the quantification of tissue fat content [6142-52]Walter, D. J. / Tkaczyk, E. J. / Wu, X. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Dual-energy imaging using a photon counting detector with electronic spectrum-splitting [6142-53]Bornefalk, H. / Lundqvist, M. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
X-ray tube voltage and image quality in adult and pediatric CT [6142-94]Huda, W. / Ogden, K. M. / Scalzetti, E. M. / Lavallee, R. L. / Samei, E. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Novel x-ray image sensor using CsBr:Eu phosphor for computed radiography [6142-104]Nanto, H. / Takei, Y. / Nishimura, A. / Nakano, Y. / Shouji, T. / Yanagita, T. / Kasai, S. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
A Monte Carlo investigation on the impact of scattered radiation on mammographic resolution and noise [6142-118]Saunders, R. S. / Samei, E. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Response of a Csl/amorphous-Si flat panel detector as function of incident x-ray angle [6142-123]Tkaczyk, J. E. / Claus, B. / Trotter, D. G. / Eberhard, J. W. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
A method of accuracy evaluation of line spread function (LSF) and point spread function (PSF) measured in computed tomography [6142-126]Wada, S. / Ohkubo, M. / Matsumoto, T. / Nishizawa, K. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Fat/water separation in a single MRI image with arbitrary phase shift [6142-142]Tisdall, M. D. / Atkins, M. S. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Noise reduction for low-dose helical CT by 3D penalized weighted least-squares sinogram smoothing [6142-159]Wang, J. / Li, T. / Lu, H. / Liang, Z. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Computer simulation of FDK reconstruction with the in-line holographic projection data [6142-161]Cai, W. / Ning, R. / Yang, D. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Measurement of slice thickness and in-plane resolution on radiographic tomosynthesis system using modulation transfer function (MTF) [6142-195]Li, B. / Saunders, R. / Uppaluri, R. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Quantification for contrast-enhanced digital breast tomosynthesis [6142-13]Carton, A.-K. / Li, J. / Albert, M. / Chen, S. / Maidment, A. D. A. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
High voltage protection in active matrix flat-panel imagers [6142-28]Lehnert, J. / Zhao, W. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Scatter correction for cone-beam computed tomography using simulated object models [6142-48]Bertram, M. / Wiegert, J. / Rose, G. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Breast cancer diagnosis using neutron stimulated emission computed tomography: dose and count requirements [6142-60]Floyd, C. E. / Bender, J. E. / Harrawood, B. / Sharma, A. C. / Kapadia, A. / Tourassi, G. D. / Lo, J. Y. / Howell, C. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Image quality improvement and exposure dose reduction with the combined use of x-ray modulation and Boost3D [6142-88]Kazama, M. / Tsukagoshi, S. / Okumura, M. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Dual-energy contrast-enhanced digital mammography (DE-CEDM): optimization on digital subtraction with practical x-ray low/high-energy spectra [6142-95]Chen, B. / Jing, Z. / Smith, A. P. / Parikh, S. / Parisky, Y. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Novel NPS measurement method for medical liquid crystal display using periodic components subtraction technique [6142-125]Ichikawa, K. / Kodera, Y. / Horii, A. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Image reconstruction in cone-beam CT with a spherical detector using the BPF algorithm [6142-164]Zuo, N. / Zou, Y. / Jiang, T. / Pan, X. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Effect of random coincidences for quantitative cardiac PET studies using 3D oxygen-15 water scans [6142-165]Bouchareb, Y. / Thielemans, K. / Spinks, T. / Rimoldi, O. / Camici, P. G. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Comparison of full-scan and half-scan for cone beam breast CT imaging [6142-167]Chen, L. / Shaw, C. C. / Lai, C. / Altunbas, M. C. / Wang, T. / Tu, S. / Liu, X. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Radial intensity projection for lumen: application to CT angiographic imaging [6142-172]Kokubun, H. / Miyazaki, O. / Hayashi, H. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Simulated and experimental technique optimization of dual-energy radiography: abdominal imaging applications [6142-55]Sabol, J. M. / Wheeldon, S. J. / Jabri, K. N. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Optimal spectra for indirect detector breast tomosynthesis [6142-57]Glick, S. J. / Gong, X. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
An iteration algorithm in dual-energy x-ray imaging based on polychromatic physics model [6142-99]Tang, S. / Mou, X. / Luo, T. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Dynamic platform for moving organ imaging [6142-120]Grosjean, R. / Guerra, R. / Lorentz, C. / Pasquier, C. / Vuissoz, P.-A. / Claudon, M. / Felblinger, J. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Stationary x-ray source for tomographic medical imaging (Cum Laude Poster Award) [6142-132]Schwoebel, P. R. / Boone, J. M. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
An inversion method for the cone-beam transform (Special Image Reconstruction Award) [6142-157]Yarman, C. E. / Yazici, B. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Interpolation method for the Mojette transform [6142-163]Servieres, M. / Normand, N. / Guedon, J. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Optimized time window for cardiac CT reconstruction based on Doppler tissue imaging (DTI) data [6142-20]Guerra, R. / Codreanu, A. / Ponvianne, Y. / Grosjean, R. / Felblinger, J. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
ECG gated continuous circular cone-beam multi-cycle reconstruction for in-stent coronary artery imaging: a phantom study [6142-21]van Stevendaal, U. / Koken, P. / Begemann, P. G. C. / Koester, R. / Adam, G. / Grass, M. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
High-performance dual-energy imaging with a flat-panel detector: imaging physics from blackboard to benchtop to bedside [6142-50]Siewerdsen, J. H. / Shkumat, N. A. / Dhanantwari, A. C. / Williams, D. B. / Richard, S. / Daly, M. J. / Paul, N. S. / Moseley, D. J. / Jaffray, D. A. / Yorkston, J. et al. | 2006
-
Effect of ghosting on the modulation transfer function of amorphous selenium based flat panel detectors [6142-64]Chowdhury, L. / Tousignant, O. / DeCrescenzo, G. / Gauthier, P. / Leboeuf, J. / Rowlands, J. A. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
A practical method for measuring the H matrix of digital x-ray and cone beam CT imaging systems [6142-66]Kyprianou, I. S. / Badano, A. / Gallas, B. D. / Park, S. / Myers, K. J. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Temperature coefficients and noise performance and studies for the back-illuminated arrays for medical imaging applications [6142-103]Goushcha, A. O. / Tabbert, B. / Goushcha, I. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Transparent BaCl~2Eu^2^+glass-ceramic scintillator [6142-105]Chen, G. / Johnson, J. / Schweizer, S. / Woodford, J. / Newman, P. / MacFarlane, D. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Photo-bleaching compensation for autofocus algorithms in fluorescence microscope applications [6142-137]Soda, P. / Iannello, G. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Combination of CT scanning and fluoroscopy imaging on a flat-panel CT scanner [6142-91]Grasruck, M. / Gupta, R. / Reichardt, B. / Suess, C. / Schmidt, B. / Stierstorfer, K. / Popescu, S. / Brady, T. / Flohr, T. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Dual-energy imaging using a digital scanned multi-slit system for mammography: evaluation of a differential beam filtering technique [6142-96]Bornefalk, H. / Hemmendorff, M. / Hjarn, T. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Choosing the right resolution for vessel visualization in MR angiography [6142-130]Vemuri, P. / Kholmovski, E. G. / Parker, D. L. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Fast parallel MRI reconstruction using B-spline approximation (PROBER) [6142-134]Petr, J. / Kybic, J. / Hlavac, V. / Muller, S. / Bock, M. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
First attempt of the medical application of the refraction-based computed tomography [6142-135]Maksimenko, A. / Hashimoto, E. / Ando, M. / Sugiyama, H. / Yuasa, T. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
A new approach of 3D SPECT reconstruction for near-field coded aperture imaging [6142-158]Hong, B. / Mu, Z. / Liu, Y.-H. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
An approximate cone beam reconstruction algorithm for gantry-tilted CT [6142-175]Yan, M. / Zhang, C. / Liang, H. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Low-contrast lesion detection in tomosynthetic breast imaging using a realistic breast phantom [6142-192]Zhou, L. / Oldan, J. / Fisher, P. / Gindi, G. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Optimization of operating conditions in photon-counting multi-slit mammography based on Si-strip detectors [6142-10]Aslund, M. / Cederstrom, B. / Lundqvist, M. / Danielsson, M. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Toward time resolved 4D cardiac CT imaging with patient dose reduction: estimating the global heart motion [6142-19]Taguchi, K. / Segars, W. P. / Fung, G. S. K. / Tsui, B. M. W. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Reducing temporal fluctuations in MRI with the multi-channel method SENSE [6142-25]Moeller, S. / Van de Moortele, P.-F. / Goerke, U. / Ugurbil, K. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Description of statistical theory of magnetic resonance imaging (MRI) [6142-26]Lei, T. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Single-photon spatial and energy resolution enhancement of a columnar CsI(TI)/EMCCD gamma-camera using maximum-likelihood estimation [6142-65]Miller, B. W. / Barber, H. B. / Barrett, H. H. / Shestakova, I. / Singh, B. / Nagarkar, V. V. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
A comparison of the performance of new screen-film and digital mammography systems [6142-119]Monnin, P. / Gutierrez, D. / Castella, C. / Lepori, D. / Verdun, F. R. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Anode heel affect in thoracic radiology: a visual grading analysis [6142-121]Mearon, T. / Brennan, P. C. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Reconstruction implementation based on a flat panel detector cone beam breast CT [6142-162]Yang, D. / Ning, R. / Conover, D. / Yu, Y. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
A backprojection filtered image reconstruction algorithm for circular cone-beam CT [6142-168]Li, L. / Chen, Z. / Zhang, L. / Xing, Y. / Kang, K. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Evaluation of an algorithm for simultaneous reconstruction of activity and attenuation in 3D PET [6142-173]Soto, J. L. P. / Furuie, S. S. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Cone-beam CT reconstruction using a nonlinear weighted filtered backprojection from half-scan data [6142-179]Gomi, T. / Koshida, K. / Miyati, T. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Preliminary assessment of the temporal subtraction of tomosynthesis images for improved detection of pulmonary nodules [6142-194]Li, C. M. / Dobbins, J. T. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Evaluation of a photon-counting breast tomosynthesis imaging system [6142-11]Maidment, A. D. A. / Ullberg, C. / Lindman, K. / Adelow, L. / Egerstrom, J. / Eklund, M. / Francke, T. / Jordung, U. / Kristoffersson, T. / Lindqvist, L. et al. | 2006
-
Design considerations in cardiac CT [6142-17]Bruder, H. / Stierstorfer, K. / McCullough, C. / Raupach, R. / Petersilka, M. / Grasruck, M. / Suess, C. / Ohnesorge, B. / Flohr, T. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Effect of heart rate on CT angiography using the enhanced cardiac model of the 4D NCAT [6142-18]Segars, W. P. / Taguchi, K. / Fung, G. S. K. / Fishman, E. K. / Tsui, B. M. W. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Low-noise pixel architecture for advanced diagnostic medical x-ray imaging applications [6142-29]Izadi, M. H. / Karim, K. S. / Nathan, A. / Rowlands, J. A. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Three-dimensional columnar CsI model for x-ray imaging system simulations using MANTIS: validating for noise, blur, and light output [6142-32]Badano, A. / Kyprianou, I. S. / Sempau, J. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Development of next generation digital flat panel catheterization system: design principles and validation methodology [6142-49]Belanger, B. / Betraoui, F. / Dhawale, P. / Gopinath, P. / Tegzes, P. / Vagvolgyi, B. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
CMOS cassette for digital upgrade of film-based mammography systems [6142-62]Baysal, M. A. / Toker, E. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Experimental comparison of cone beam CT (CBCT) reconstruction and multiview reconstruction (MVR) for microangiography (MA) detector system [6142-85]Patel, V. / Kuhls, A. T. / Noel, P. B. / Walczak, A. / Ionita, C. N. / Chityala, R. / Tranquebar, R. / Rangwala, H. S. / Kasodekar, S. S. / Hoffmann, K. R. et al. | 2006
-
Flat panel detector-based cone beam CT for dynamic imaging: system evaluation [6142-84]Ning, R. / Conover, D. / Yu, Y. / Zhang, Y. / Cai, W. / Yang, D. / Lu, X. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Impact of CT detector pixel-to-pixel crosstalk on image quality [6142-87]Engel, K. J. / Spies, L. / Vogtmeier, G. / Luhta, R. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Nonlinear dual-spectral image fusion for improving cone-beam-CT-based breast cancer diagnosis [6142-97]Chen, Z. / Ning, R. / Conover, D. / Willison, K. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
A robust x-ray tube spectra measuring method by attenuation data [6142-129]Yang, Y. / Mou, X. / Chen, X. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Novel SNR determination method in parallel MRI [6142-133]Miyati, T. / Imai, H. / Ogura, A. / Doi, T. / Tsuchihashi, T. / Machida, Y. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
MR image reconstruction using the GPU [6142-138]Schiwietz, T. / Chang, T. / Speier, P. / Westermann, R. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Spectral imaging of skin: experimental observations and analyses [6142-140]Kerekes, J. / Subramanian, N. / Kearney, K. / Schad, N. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Elliptical extrapolation of truncated 2D CT projections using Helgason-Ludwig consistency conditions [6142-156]Van Gompel, G. / Defrise, M. / Van Dyck, D. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Accelerating 3D cone beam T-FDK algorithm on commodity PC graphics hardware [6142-170]Dai, Z. / Chen, Z. / Xing, Y. / Zhang, L. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Impact of noise-on-image reconstruction for diffuse optical tomography [6142-22]Nielsen, T. / Kohler, T. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Generalized auto-calibrating technique for image reconstruction from sensitivity encoded MRI data [6142-24]Kholmovski, E. G. / Parker, D. L. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Experimental spectral measurements of heavy K-edge filtered beams for x-ray computed mammotomography [6142-67]Crotty, D. J. / McKinley, R. L. / Tornai, M. P. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Comparison measurements of DQE for two flat panel detectors: fluoroscopic detector vs. cone beam CT detector [6142-92]Benitez, R. B. / Ning, R. / Conover, D. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Substrate effect on indirect digital radiography system performance [6142-127]Lubinsky, A. R. / Yip, K. L. / Trauernicht, D. P. / Yorkston, J. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Acceleration of fluoro-CT reconstruction for a mobile C-arm on GPU and FPGA hardware: a simulation study [6142-166]Xue, X. / Cheryauka, A. / Tubbs, D. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
A new 2D-tiled detector for multislice CT [6142-30]Luhta, R. / Chappo, M. / Harwood, B. / Mattson, R. / Salk, D. / Vrettos, C. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Light transport in trabecular bone: Monte Carlo simulation based on 3D triangle meshes [6142-56]Margallo-Balbas, E. / French, P. J. / Wieringa, P. A. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Theoretical investigation of very high-quantum-efficiency segmented crystalline detectors for low-contrast visualization in megavoltage cone-beam CT [6142-61]Wang, Y. / Antonuk, L. E. / El-Mohri, Y. / Sawant, A. / Zhao, Q. / Du, H. / Li, Y. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
New light-amplifier-based detector designs for high spatial resolution and high sensitivity CBCT mammography and fluoroscopy [6142-63]Rudin, S. / Kuhls, A. T. / Yadava, G. K. / Josan, G. C. / Wu, Y. / Chityala, R. N. / Rangwala, H. S. / Ionita, N. C. / Hoffmann, K. R. / Bednarek, D. R. et al. | 2006
-
Geometry calibration phantom design for 3D imaging [6142-86]Claus, B. E. H. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Amplitude correlated 4-dimensional cone beam CT [6142-90]Lu, J. / Zhu, X. R. / Munro, P. / Chi, P. M. / Mohan, R. / Pan, T. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
A computation method of dual-energy x-ray imaging [6142-100]Mou, X. / Tang, S. / Hong, W. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Energy-dependent scintillation intensity of fluorozirconate-based glass-ceramic x-ray detectors [6142-106]Schweizer, S. / Henke, B. / Koneke, S. / Johnson, J. A. / Chen, G. / Woodford, J. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Spectral imaging of near-surface oxygen saturation [6142-143]Subramanian, N. R. / Kerekes, J. P. / Kearney, K. / Schad, N. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Optimizing filtered backprojection reconstruction for a breast tomosynthesis prototype device [6142-15]Mertelmeier, T. / Orman, J. / Haerer, W. / Dudam, M. K. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Motion artifacts from an inverse-geometry CT system with multiple detector arrays [6142-16]Mazin, S. / Pelc, N. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Comparison of three sinogram restoration methods [6142-34]Forthmann, P. / Kohier, T. / Defrise, M. / La Riviere, P. J. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Evaluation of a physical based approach of scattered radiation correction in cone beam CT with an anthropomorphic thorax phantom [6142-47]Rinkel, J. / Gerfault, L. / Esteve, F. / Dinten, J. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
The impact of cardiac gating on the detection of coronary calcifications in dual-energy chest radiography: a phantom study [6142-51]Sabol, J. M. / Liu, R. / Saunders, R. / Markley, J. / Moreno, N. / Seamans, J. / Wiese, S. / Jabri, K. / Gilkeson, R. C. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Monte Carlo simulation of x-ray scatter based on patient model from digital breast tomosynthesis [6142-59]Liu, B. / Wu, T. / Moore, R. H. / Kopans, D. B. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
The line-noise-reduction for low dose x-ray fluoroscopy with the flat panel detector [6142-101]Nakamura, T. / Ikeda, S. / Suzuki, K. / Takenouchi, S. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Dosimetric and image quality assessment of different acquisition protocols of a novel 64-slice CT scanner [6142-124]Vite, C. / Mangini, M. / Strocchi, S. / Novario, R. / Tanzi, F. / Carrafiello, G. / Conte, L. / Fugazzola, C. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
An inversion method for the exponential Radon transform based on the harmonic analysis of the Euclidean motion group [6142-155]Yarman, C. E. / Yazici, B. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
High-speed large angle mammography tomosynthesis system [6142-12]Eberhard, J. W. / Staudinger, P. / Smolenski, J. / Ding, J. / Schmitz, A. / McCoy, J. / Rumsey, M. / Al-Khalidy, A. / Ross, W. / Landberg, C. E. et al. | 2006
-
High-resolution multiphoton optical tomography of tissues: an in vitro and in vivo study [6142-23]Riemann, I. / Schenke-Layland, K. / Ehlers, A. / Dimitrow, E. / Kaatz, M. / Elsner, P. / Martin, S. / Konig, K. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Performance evaluation of dual-crystal APD-based detector modules for positron emission tomography [6142-27]Pepin, C. M. / Berard, P. / Cadorette, J. / Tetrault, M.-A. / Leroux, J.-D. / Michaud, J.-B. / Robert, S. / Dautet, H. / Davies, M. / Fontaine, R. et al. | 2006
-
Development of a computer-generated model for the coronary arterial tree based on multislice CT and morphometric data [6142-58]Fung, G. S. K. / Segars, W. P. / Taguchi, K. / Fishman, E. K. / Tsui, B. M. W. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
A method to measure the temporal MTF to determine the DQE of fluoroscopy systems [6142-69]Friedman, S. N. / Cunningham, I. A. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
A novel method to characterize the MTF in 3D for computed mammotomography [6142-70]Madhav, P. / McKinley, R. L. / Samei, E. / Bowsher, J. E. / Tornai, M. P. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
A new scanning device in CT with dose reduction potential [6142-93]Tischenko, O. / Xu, Y. / Hoeschen, C. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Light induced degradation in amorphous silicon photodiodes and implication for diagnostic medical imaging application [6142-102]Khodami, I. / Adachi, M. / Malhotra, M. / Taghibakhsh, F. / Rowlands, J. / Karim, K. / Kavanagh, K. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Quarantine MAP reconstruction of PET/CT data using dual priors [6142-160]Chan, C. / Meikle, S. / Fulton, R. / Cai, T. W. / Feng, D. D. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Implementation of strip-area system model for fan-beam collimator SPECT reconstruction [6142-169]Ye, H. / Krol, A. / Feiglin, D. H. / Lipson, E. D. / Lee, W. / Coman, I. L. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Image reconstruction from discrete Chebyshev moments via formation of lookup tables [6142-174]Bayraktar, B. / Bernas, T. / Robinson, J. P. / Rajwa, B. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Implementing an iterative reconstruction algorithm for digital breast tomosynthesis on graphics processing hardware [6142-176]Goddard, I. / Wu, T. / Thieret, S. / Berman, A. / Bartsch, H. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
A correction method for unfunctional cell distortion using orthogonal polynomials [6142-177]Tang, S. / Mou, X. / Yan, H. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Implementation of a spiral CT backprojection algorithm on the cell broadband engine processor [6142-178]Bockenbach, O. / Goddard, I. / Schuberth, S. / Seebass, M. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Preliminary validation of a new methodology for estimating dose reduction protocols in neonatal chest computed radiographs [6142-68]Don, S. / Whiting, B. R. / Hildebolt, C. F. / Sehnert, W. J. / Ellinwood, J. S. / Topfer, K. / Masoumzadeh, P. / Kraus, R. A. / Kronemer, K. A. / Herman, T. et al. | 2006
-
A new reconstruction algorithm for Radon data [6142-82]Xu, Y. / Tischenko, O. / Hoeschen, C. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Data pre-processing for quantification in tomography and radiography with a digital flat panel detector [6142-83]Rinkel, J. / Gerfault, L. / Esteve, F. / Dinten, J.-M. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Windmill artifacts analysis in MSCT [6142-89]Amir, O. / Sabo-Napadensky, I. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
New microangiography system development providing improved small vessel imaging, increased contrast-to-noise ratios, and multiview 3D reconstructions [6142-131]Kuhls, A. T. / Patel, V. / Ionita, C. / Noel, P. B. / Walczak, A. M. / Rangwala, H. S. / Hoffmann, K. R. / Rudin, S. / SPIE-- the International Society for Optical Engineering et al. | 2006
-
Selective photon counter for digital x-ray mammography tomosynthesis (Honorable Mention Poster Award) [6142-193]Goldan, A. H. / Karim, K. S. / Rowlands, J. A. / SPIE-- the International Society for Optical Engineering et al. | 2006